A: When different adapter is used, since the voltages are different, it will burn your unit. It is very crucial that you use our adapters for our products only. Please contact us immediately
A: Pins corrode due to too much sterilization. Any liquid left from sterilization will cause corrosion.
Make sure to have the unit dry after sterilization.
A: This is the most common complaint that we receive from the customers.
Almost 98% cases are due to lack of charging.
Please keep the unit charged at ALL TIMES for longer battery lifespan.
A: The heating stops after 3 seconds and activates again after a short pause because we have upgraded our software to prevent any burns with longer heating cycle during the treatment for patient safety.
A: “Er1” shows on LED display when the heater is in a bad contact with the PCB inside. Hold the Gutta-Easy Gun Body steady while rotating the heating end at least 360° to realign the PCB's contact with the heater. If the device still displays "Er1", please request for an inspection.
"Low" show on LED display when there is no battery, please charge the battery.
A: If the gun was not charged, charge it, and try turning it on again.
Also, try the reset button on the bottom of the handpiece. Press the square button. Once pressed, it will click in. Leave it for few seconds and press it again. Unit will turn on. If it still doesn’t turn on, please request for an inspection.
A: Please request for an inspection. Our repair department would have to disassemble the unit and remove the pellet inside.
To prevent this from happening, please use 1 pellet at a time. Do not use 2. Once the pellet is inserted, insert the plunger pushing towards the mouth of the handpiece.
Also, make sure the plunger has the white o-ring (Teflon Ring) in place. Without Teflon Ring, Gutta Percha Pellet can move incorrectly and get stuck inside device.
A: “OHE” appears when the temperature is set lower than the current set temperature.
A: LED display may flicker due to inconsistent flow of electricity via adapters. Check your adapter connection on the back of the unit and the power outlets on the wall.
1. The device should be cleaned with a soft cloth with alcohol or ethyl alcohol and then with a dry cloth.
2. Do not use solvent like thinner/benzene, which may damage the housing.
3. Do not use ultrasonic sterilization, self-sterilization in home, dry sterilization, cresol, and infrared light. These methods can cause damage to the device.
A: LED display may flicker due to inconsistent flow of electricity via adapters. Check your adapter connection on the back of the unit and the power outlets on the wall.
A: If any button is pressed during initialization, make sure the buttons are not being held down. If the problem persists, please contact our customer service department for further assistance.
A: Left Sensor / Right Sensor malfunction. If the problem persists, please contact our customer service department for further assistance.
A: Device is heated to over 73°C (163°F) and forced off until the temperature drops to set temperature. If the problem persists, please contact our customer service department for further assistance.
1. The device should be cleaned with a soft cloth with alcohol or ethyl alcohol and then with a dry cloth.
2. Do not use solvent like thinner/benzene, which may damage the housing.
3. Do not use ultrasonic sterilization, self-sterilization in home, dry sterilization, cresol, and infrared light. These methods can cause damage to the device.
A: After connecting the D6 head, there's a white rubber connector just above the arms. Tighten the screws on the connector with wrench connected in the head set box '
A: When different adapter is used, since the voltages are different, it will burn your unit. It is very crucial that you use our adapters for our products only. Please contact us immediately
A: Pins corrode due to too much sterilization. Any liquid left from sterilization will cause corrosion.
Make sure to have the unit dry after sterilization.
Cleaning and Disinfection Instructions must be followed after each use.
Suggested Surface Disinfectant:
Do NOT Use:
Disassembly and Inspection:
Handpiece and Charger Base
Cleaning:
Disinfecting:
Drying:
Light Guide (Fiber Optic Glass)
Cleaning:
Rinsing:
Sterilizing:
Package in an FDA cleared wrap prior to sterilization. Sterilization can be performed in gravity autoclave at 132° C / 270° F for 15 minutes with a 15-30 minute drying time.
Cleaning:
Rinsing:
Sterilizing:
Package in a FDA cleared wrap prior to sterilization. Sterilization can be performed in a gravity autoclave at 132°C / 270°F for 15 mins. with a 15-30 min. drying time.
Reassembly & Storage