Shopify logo loading animation
Skip to main content

Free Shipping for $100+ (U.S.) ㅤ 📞 1-310-328-1600
Search
My Account
Wishlist
Dentazon Corp. ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ ‎ DENTAZON CORP.

Menu

  • Home
  • About Us
  • Products
    • Products 1

      • Anesthetic / Composite Equipment
      • Orthodontic LED Curing Light / Stripping bur
      • Endodontic Obturation System
      • - Endodontic Obturation Parts
      • Ultrasonic Irrigation
      • Endodontic Root Canal File
      • EZ Bite Block / EZ Blade Remover

      Products 2

      • Mixing Tip / EZ VPS Connector / Bite Tray / Intraoral tip
      • Impression Cartridge
      • Ooze-Free Syringe
      • Tooth Bleaching System
      • Intraoral Camera
      • Operatory Light
      • Auto Mixing Magic Hair Dye

  • Service Request
  • Events
    • CDA Anaheim 2026

      07 April

      CDA Anaheim 2026

      IDEX 2026

      24 March

      IDEX 2026

      CDS Midwinter 2026

      16 December

      CDS Midwinter 2026

  • Contact Us
  • Sale
      • APRIL SALE
  • mobile menu not delete
My Cart: 0 item(s)

It appears that your cart is currently empty

  • Home
  • All Products
  • Cybird™ Gold : Dual Band LED Curing Light
Cybird gold dual band LED curing light
Cybird gold dual band LED curing light with charger base.
45 and 90 degree curing light light guides with transilluminator tip.
Sale Off
Cybird gold dual band LED curing light
Cybird gold dual band LED curing light
Cybird gold dual band LED curing light with charger base.
45 and 90 degree curing light light guides with transilluminator tip.

Cybird™ Gold : Dual Band LED Curing Light

SKU:CGD-9000

$750.00$1,190.00

Rx Only – For use by licensed dental professionals only.

  • Features

  • Specification

  • FAQ


  • Tech Profile


  • Download


  • Resources

Cybird Gold curing light

Multiwave LED Curing Light 

Cybird Gold has ergonomic compact design that fits perfectly in hands for everyone. It features NEW Dual Band LED which is compatible for most resins, improving inconvenience of carrying 2-3 curing lights for each different type of resin. 

Cybird Gold generates low heat and still maintains high intensity for efficient curing. Also, Cybird Gold helps doctors to reduce chair time caused during the treatment with 1 short polymerization cycle. 

90 degree light guide, 45 degree light guide, transilluminator tip

3 complimentary light guides are included

(Transilluminator tip, 45°, 90° light guides)


3 Curing Modes

  • High Power Mode : 1,500 mW/cm² (5, 10, 15, 20 sec.)
  • Plasma Emulation Mode : 2,500 mW/cm² (2, 3, 4 sec.)
  • Ortho Mode : Easier for orthodontic bonding (16 times of 3 sec. at 2,500 mW/cm²)  

Gold High Power Mode Intensity (1,500mW/cm²)

graph showing 1,500mw/cm2 of high  power mode

Appropriate for most applications

- Restorations for resins, liners, sealant, and bases

- Indirect restorations such as veneers, inlays/onlays 

and crowns

Gold Plasma Mode Intensity (2,500mW/cm²)

Graph of 2,500 mw/cm2 in plasma mode

Emulated plasma arc equivalent light intensity for 

rapid curing and orthodontic bonding

- 2 seconds cure for most 2mm depth composites

- Orthodontic bonding of ceramic/metal brackets, 

tubes, and bands 


Design Upgrades

cyborg gold curing light with numbering 1,2,3 on the light guide, button, and battery

1. 8mm fiber-optic light guide

The premium low profile fiber-optic light guide greatly improves the Gold's light emission and access for posterior procedures

2. Intuitive button control / LED Display

The newly improved silicone dual control button allows to change the device settings and precisely displays various conditions of the Cybird Gold such as battery level indication, mode selection and time duration

3. Enhanced battery module

A high performance lithium-ion cell is incorporated into our special 

medical-grade battery module guarantees extended battery life


New Features

9-chip led
dual band wavelength
curing light chart from 1 to 4th gerneration

9-Chip LED

Maintains stable light output & minimizes heat generation, improving LED circuit

Wavelength

More compatibility with most resins (400nm~490nm)

Lower Temperature

Generates low heat & still maintains high intensity for efficient curing


Related Products

Sale Off
Cybird XD plasma emulation LED curing light.

New Cybird™ XD : Plasma Emulation LED Curing Light

$850.00$550.00
Sale Off
Cybird Ultra

[COMING SOON] Cybird™ Ultra LED Curing Light

$0.00
Sale Off
Single-Sided - IPR Metal Auto Interproximal Reduction Strip

Single-Sided - IPR Metal Auto Interproximal Reduction Strip

$17.99
Sale Off
Double Sided - IPR Metal Auto Interproximal Reduction Strip

Double Sided - IPR Metal Auto Interproximal Reduction Strip

$17.99

Specification

Wavelength

400 ~ 490nm

Irradiance Intensity 

High Power Mode = 1,500 mW/cm² (5,10,15,20 sec.)

Plasma Emulation Mode = 2,500 mW/cm² (2,3,4 sec.)

Ortho Mode (16 times of 3 sec. at 2,500 mW/cm²) 

Dimensions 

Handpiece = 150 x 30 x 22.6 (mm)

Charger Base = 120 x 120 x 85 (mm)

Charging Time 

2 hours

Operation Time 

High Power Mode, 10 sec. x 300 (when fully charged) 

FAQS

Charger has a burning smell

When different adapter is used, since the voltages are different, it will burn your unit. It is very crucial that you use our adapters for our products only. 
Please contact us immediately.

Battery pins / Charger pins are corroded and will not charge    

Pins corrode due to too much sterilization.  Any liquid left from sterilization will cause corrosion.

Make sure to have the unit dry after sterilization.

Cleaning Instructions   

Cleaning and Disinfection Instructions must be followed after each use.


Suggested Surface Disinfectant:


- CaviCide®(or equivalent quaternary/alcohol blend disinfectant)


Do NOT Use:


- Glutaraldehyde

- Denatured alcohol

- Lysol®

- Phenol or phenolic cleaner

 -Ammonia complex

 -Iodine complex solutions


Disassembly and Inspection:


1. Remove elastomer rubber light shield from glass fiber-optic light guide.

2. Remove light guide from handpiece by gently pulling it straight out. Examine light guide ends for damage or composite material adhering to the tip. Examine light shield for tears or distortion. Replace if any damage is noted on either component.


Handpiece and Charger Base


Cleaning:


1. Initial cleaning of handpiece must begin immediately after use to prevent drying of soil and contaminants in and on the device

2. All exterior surfaces of the handpiece and charger base may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Do not allow cleaning solution to invade the interior of these components since this may adversely affect the electronics.


Disinfecting:


1. All exterior surfaces of the handpiece or charger base must be wiped and wetted with CaviWipes® or a cloth soaked in surface disinfectant.

2. Allow the CaviCide® surface disinfectant to reside of the surface for a minimum of 3 minutes. Do not allow the disinfectant to dry on the surface.


Drying:


1. Use a clean dry cloth to dry the exterior surfaces. Do not allow fluids to accumulate in the charger base socket as this may adversely affect the electronics.


Light Guide (Fiber Optic Glass)


Cleaning:


1. Initial cleaning of the fiber optic light guide must begin immediately after use to prevent drying of soil and contaminants in and on the device.

2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Use a soft brush to remove contaminants in the junction between the glass rod and metal ferrule if necessary.

3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time 10 minutes.


Rinsing:


1. Using clean tap water on a clean cloth, wipe away residual disinfectant


Sterilizing:


Package in an FDA cleared wrap prior to sterilization. Sterilization can be performed in gravity autoclave at 132° C / 270° F for 15 minutes with a 15-30 minute drying time.


Light Shield (Elastomer Rubber)


Cleaning:


1. Initial cleaning of the light shield must begin immediately after use to prevent drying of soil and contaminants in and on the device.

2. All exterior surfaces may be wiped with CaviWipes® or a cloth soaked in surface disinfectant to remove gross soil. Examine the light shield for damage and discard if any cuts, tears, or distortion is noted.

3. Clean thoroughly using an ultrasonic cleaner. Minimum cycle time is 10 mins.

4. The elastomer rubber light shield can be cleaned, sterilized and reused up to 5 times, discard after that.


Rinsing:


1. Using clean tap water on a clean cloth, wipe away residual disinfectant.


Sterilizing:


Package in a FDA cleared wrap prior to sterilization. Sterilization can be performed in a gravity autoclave at 132°C / 270°F for 15 mins. with a 15-30 min. drying time.


Reassembly & Storage


1. After all parts have been allowed to dry completely, gently insert the metal end of the glass fiber optic light guide into the circular opening of the handpiece. Make sure the light guide is completely inserted until it stops. This will position the light guide properly with the internal LED light source.

2. Insert an elastomer rubber light shield over the end of the fiber optic light guide.

3. Store the reassembled Cybird Curing light in a clean, dry location, preferably on the Cybird charger base so that the battery will be completely charged prior to the next use.

Inside Dentistry Tech Profile

Article of In proctice Magazine on Cybird Gold Curing Light. The diverse and powerful fourth-generation curing light

This “Dual Band LED Curing Light” Cybird Gold is specially developed to polymerize a wide range of materials. 


This fourth-generation curing light fully adjusts to different types of dental treatments and can help reduce the expense or time associate with Treatment by offering one short polymerization cycle. 


The Cybird Gold’s compact and ergonomic design fits perfectly in the hand, and the Dual-Band LED makes it compatible with most resins.


Inside Dentistry, April 2016 

Published Dental Studies and Case Reports Involving DXM / Dentazon Products

High-Irradiance Light Curing, Depth of Cure, and Pulpal Temperature

Study: Effect of High-Irradiance Light Curing on Exposure Times and Pulpal Temperature of Adequately Polymerized Composite

 

Authors: 

William E Slack, Advanced Education in General Dentistry

Erin M Yancey, Advanced Education in General Dentistry

Wen Lien, USAF Dental Research and Consultation Service

Ryan Sheridan, USAF Prosthodontics Residency

Rodney Phoenix, Postgraduate Dental College, Uniformed Services University of the Heatlh Sciences

Kraig Vandewalle, Advanced Education in General Dentistry Residency, Uniformed Services University of the Health Sciences

 

Journal: Dental Materials Journal 2020; 39(6): 976–983

 

Publication Date: December 2020

 

Product Used: Cybird™ XD

 

Summary: This in vitro study compared high-irradiance curing lights, including Cybird™ XD, Valo, S.P.E.C. 3 LED, and Flashmax P3. The study found that the manufacturer-recommended short exposure times did not adequately polymerize the tested composite at a 2 mm depth. Under the study protocol, Cybird™ XD achieved adequate polymerization at 2 mm after 9 seconds. The pulpal temperature increase at this experimental exposure time remained below the threshold associated with possible pulpal injury.

 

Result Highlight: Cybird™ XD achieved adequate 2 mm polymerization at 9 seconds in this study, with pulpal temperature increase remaining below the cited safety threshold.

 

Go to Study →

Dental Radiometer Accuracy Study

Study: On the Inaccuracies of Dental Radiometers

 

Authors: 

Adrian C Shortall, College of Medical and Dental Sciences, Institute of Clinical Studies, University of Birmingham, Birmingham, United Kingdom

Mohammed A Hadis, College of Medical and Dental Sciences, Institute of Clinical Studies, University of Birmingham, Birmingham, United Kingdom

William M Palin, College of Medical and Dental Sciences, Institute of Clinical Studies, University of Birmingham, Birmingham, United Kingdom

Editor: Fabian Huettig, Eberhard-Karls-Universitat Tubingen Medizinische Fakultat, GERMANY

 

Journal: Public Library of Science ONE (PLOS ONE) 16(1): e0245830. https://doi.org/10.1371/journal.pone.0245830

 

Publication Date: January 2021

 

Product Used: Cybird™ XD

 

Summary: This study evaluated the accuracy of commercial dental radiometers when measuring LED light-curing units, including Cybird™ XD. The authors reported substantial discrepancies between radiometer readings and gold-standard integrating sphere measurements, showing that dental radiometers may not reliably estimate actual light output for all curing lights.

 

Result Highlight: The study supports the importance of accurate light-output measurement methods when evaluating curing-light performance.

 

Go to Study →

Translucency Testing of Zirconia and Hybrid Ceramic Crowns

Study: Translucency of Zirconia and Hybrid Ceramic Monolithic Crowns

 

Authors: 

P Baldissara, Dental School, University of Bologna, Bologna, Italy

C Mazzitelli, Dental School, University of Bologna, Bologna, Italy

T Maravic, Dental School, University of Bologna, Bologna, Italy

U Josic, Dental School, University of Bologna, Bologna, Italy

V Wandscher, Dental School, University of Bologna, Bologna, Italy

L Breschi, Dental School, University of Bologna, Bologna, Italy

 

Journal: Dental Materials Journal 2023; 39(1): e12

 

Publication Date: August 2023

 

Product Used: Cybird™ Gold 

 

Summary: This dental materials study evaluated translucency of zirconia and hybrid ceramic monolithic crowns using white light and blue light transmission testing. A DXM Cybird™ Gold lamp was used for the blue-light transmission measurement. The study found differences in translucency depending on material type and sintering protocol.

 

Result Highlight: Cybird™ Gold was used as a blue-light source in translucency testing of ceramic and hybrid ceramic crown materials.

 

Go to Study →

※ Summaries are based on the cited articles. All rights to the original article belong to the authors and/or publisher. Product names and trademarks mentioned in the cited studies belong to their respective owners.

Contact

Dentazon Corp.

24414 S. Main St. Ste 203

Carson, CA 90745 U.S.                                                                                  

☎ +1-310-328-1600

whatsapp +1-310-612-6468

          

Please call us if you have trouble using our website

INFORMATION

  • ABOUT US
  • TERMS AND CONDITIONS
  • PRIVACY POLICY
  • CONTACT US
  • BLOG
  • ACCESSIBILITY STATEMENT
  • PROPOSITION 65
  • RESEARCH AND PUBLICATIONS

ACCOUNT

  • LOGIN
  • MY CART
  • MY ORDERS
  • FAQ
  • INSTRUCTIONS

NEWSLETTER

You have successfully subscribed!

SUBSCRIBE

Dentazon Corporation® 2006-2026 Dentazon.com. All Rights Reserved. Rx Only products are restricted to licensed dental professionals.